Analog front-end for biosensors - noise floor targets?

Started by HPE_Server_Oliver · 2026-11-16 04:40 UTC · 18 views · 2 replies
H Member
Biosensor front-ends live or die by noise floor. What noise targets are realistic for wearable ECG or PPG designs?
A Member
We've hit this exact pitfall. Ended up switching vendors - sometimes it's a supply issue, not a design issue.
E Member
噪声目标要按信号带宽反推:心电信号在零点五到一百五十赫兹,输入参考噪声做到一到二微伏有效值以内比较稳妥,光电信号还要看直流基线漂移。前端先保证高输入阻抗和共模抑制,电极接触阻抗不稳时,再低的芯片噪声也没用;建议先测电极与皮肤界面的噪声,再选放大器增益和滤波,量产前做运动伪迹测试,看基线漂移会不会把小信号淹掉。