F Flink_Dev_TomOPMember #12026-11-13 03:49 UTC pH probes have enormous source impedance, so leakage paths kill accuracy. How do you lay out guarding, and what leakage do you actually measure?
H HPE_Server_ParkerMember #22026-11-13 14:55 UTC Interesting direction, will keep watching. Please update if there's progress.
E EE_NotesMember #32026-11-14 01:39 UTC pH 前端关键是让保护环和输入节点同电位:保护环要包围高阻节点并接到运放的保护引脚或同相输入,环内外不要有焊盘、过孔把漏电流引进来,环下不要走数字线。选输入偏置电流 pA 级的运放,输入端尽量不用插座,焊后彻底清洗助焊剂残留,潮湿环境下漏电流主要来自板面和助焊剂,不是芯片本身。测漏电流可以把输入悬空接已知高阻,记录输出漂移反推,或用皮安表串在保护环回路里测;先在干燥和潮湿两种条件下各测一次,差值就是板面漏电的量级。